Vitenskapelig artikkel

Publisert

  • 2000

A temperature-programmed packed capillary LC method with large-volume injection on-column focusing has been developed for screening and determination of 1-(2-methoxyphenyl)piperazine derivatives of airborne toluene-2,4-diisocyanate, toluene-2,6-diisocyanate, hexamethylenediisocyanate and methylenebisphenyl-4,4-diisocyanate, based on sampling methods described in MDHS 25/3. Injection volumes up to 100 μl were successfully loaded onto the 250×0.32 mm I.D. capillary column packed with 3 μm Hypersil ODS particles. The isocyanate derivatives were loaded at 10°C and eluted by a three-step temperature program starting at 10°C for 10 min, followed by a temperature ramp of 2.5°C min−1 to 45°C and then 9.9°C min−1 to 90°C. The mobile phase consisted of acetonitrile–acetate buffer (3% triethylamine, pH 4.5) (45:55, v/v). The isocyanate derivatives were dissolved in acetonitrile–acetate buffer (3% triethylamine, pH 4.5) (30:70, v/v) to achieve sufficient focusing. The concentration limit of detection of the individual derivatives utilizing an “U” shaped flow cell with a 8.0 mm light path and an injection volume of 100 μl was 44, 87, 43 and 210 pg ml−1 for toluene-2,6-diisocyanate, hexamethylenediisocyanate, toluene-2,4-diisocyanate and methylenebisphenyl-4,4-diisocyanate, respectively. Within the investigated concentration range, 10–500 ng ml−1, the linear calibration curves gave correlation coefficients ranging from 0.994 to 0.998. The repeatability of the method with regard to retention time and [...]

Pål Molander; Karina Haugland; Geir Fladseth; Elsa Lundanes; Syvert Thorud; Yngvar Thomassen; Tyge Greibrokk
Journal of Chromatography A, 892(1-2): 67-74.
Les publikasjon