Vit. artikkel

Publisert

  • 2006

Elimination of interfering substances in urine by solid phase extraction (SPE) prior to analysis resulted in 10-fold improvement in the sensitivity of atrazine mercapturate (AM) enzyme-linked immunosorbent assay (ELISA) compared to previous reports. Of the two tested SPE systems, Oasis HLB and MCX, the mixed-mode MCX gave good recoveries (82%) of AM in spiked samples measured by ELISA, whereas the reverse-phase HLB phase was not compatible with the immunochemical method. At relatively high concentrations of urinary AM (>20 ng mL(-1)), sample dilution was effective enough for the elimination of interfering substances. The new liquid chromatography-mass spectrometry (LC-MS) method developed for AM utilizes online-SPE with Oasis HLB, column switching and a stable-isotope internal standard. The limit of quantification (0.05 ng mL(-1)) indicates improved sensitivity compared with most previously published LC-MS methods for AM. Validation of all three methods, LC-MS, ELISA+SPE and ELISA+dilution with spiked urine samples showed good correlation between the known and measured concentrations with R2 values of 0.996, 0.957 and 0.961, respectively. When a set (n=70 plus 12 blind duplicates) of urine samples from farmers exposed to atrazine was analyzed, there was a good agreement (R2=0.917) between the log normalized data obtained by ELISA+SPE and LC-MS. High correlation among...

Koivunen, Marja E.; Dettmer, Katja; Vermeulen, Roel; Bakke, Berit; Gee, Shirley J.; Hammock, Bruce D.
Analytica Chimica Acta 572(2): 180–189
Les publikasjon